In this work, we studied how the scatter-to-primary ratio changes with the energy threshold in photon counting CT. Both Monte Carlo simulations and experimental measurements demonstrated a significantly reduced SPR in the high-energy bin projection data and reduced scatter artifacts in the reconstructed CT images. A new data acquisition scheme named Mohi-Polo was proposed by pushing the high energy threshold towards maximum keV, which generates a quasi-Monochromatic high energy bin and a Polychromatic low energy bin. Utilizing the nearly scatter-free high-energy bin data, we developed an innovative solution to correct the scatter artifacts in the low-energy bin.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ran Zhang, Ke Li, and Guang-Hong Chen, "Scatter-to-primary ratio in photon counting CT," Proc. SPIE 12463, Medical Imaging 2023: Physics of Medical Imaging, 1246306 (Presented at SPIE Medical Imaging: February 20, 2023; Published: 7 April 2023); https://doi.org/10.1117/12.2654253.6323164556112.