Presentation
17 March 2023 Interchangeable imaging platform for ophthalmic robotically aligned optical coherence tomography
Author Affiliations +
Proceedings Volume PC12360, Ophthalmic Technologies XXXIII; PC123600L (2023) https://doi.org/10.1117/12.2650662
Event: SPIE BiOS, 2023, San Francisco, California, United States
Abstract
We present a flexible optical coherence tomography (OCT) imaging platform that allows for interchangeable imaging modules for specific target tissues of interest while meeting the requirements for robotically aligned OCT including integrated pupil tracking cameras. Our OCT imaging platform consisted of a fixed scan head (analogous to an SLR camera body) mounted to the robot and interchangeable anterior chamber (AC) and retinal imaging modules with their own integrated pupil cameras. We validated our system in both phantom and ex vivo porcine eyes. This flexible design enables the ability to develop new imaging modules for new robotically aligned applications.
Conference Presentation
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ryan P. McNabb, Yuan Tian, Al-Hafeez Z. Dhalla, Pablo Ortiz, Mark Draelos, Joseph A. Izatt, and Anthony N. Kuo "Interchangeable imaging platform for ophthalmic robotically aligned optical coherence tomography", Proc. SPIE PC12360, Ophthalmic Technologies XXXIII, PC123600L (17 March 2023); https://doi.org/10.1117/12.2650662