Paper
1 March 1974 Reflectance And Transmission Measurements
Philip Baumeister
Author Affiliations +
Abstract
A conventional recording spectroradiometer can be used in several ways to find the spectral reflectance and transmittance of an optical filter. Care must be taken to compensate for systematic errors introduced by the partial polarization and angular divergence of the spectroradiometer beam. Corrections must also be made for the instrumental bandwidth, spurious reflections, beam walk-off and collimation.
© (1974) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Philip Baumeister "Reflectance And Transmission Measurements", Proc. SPIE 0050, Optical Coatings: Applications and Utilization I, (1 March 1974); https://doi.org/10.1117/12.954105