25 March 1976 Effects Of Radiation Parameters On Computerized Transverse Axial Tomography
R. G. Waggener, W. D. McDavid, W. H. Payne, M. J. Dennis
Author Affiliations +
Proceedings Volume 0070, Application of Optical Instrumentation in Medicine IV; (1976) https://doi.org/10.1117/12.954615
Event: Application of Optical Instrumentation in Medicine, IV, 1975, Atlanta, United States
Abstract
Effects of kVp, voltage wave form, added filter, and tube current on three-dimensional reconstruction from x-rays have been investigated in relationship to sampling time,spectral heterogeneity, contrast, quantum noise, and patient dose. Comparison of statistical limitations with and without a water-bath indicates that scanners which employ a water-bath are comparatively inflexible in this regard. A degree of flexibility in the choice of kilovoltage and filter seems both desirable and feasible in scanners which do not use a water-bath. Minimization of patient dose is also limited by statistical factors. Reconstructions from measured data have been carried out to demonstrate the relationship between these factors.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
R. G. Waggener, W. D. McDavid, W. H. Payne, and M. J. Dennis "Effects Of Radiation Parameters On Computerized Transverse Axial Tomography", Proc. SPIE 0070, Application of Optical Instrumentation in Medicine IV, (25 March 1976); https://doi.org/10.1117/12.954615
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scanners

Computed tomography

Sensors

Copper

Radiation effects

Optical filters

Aluminum

RELATED CONTENT

Parallel biased GaAs photoconductive detector
Proceedings of SPIE (January 20 2004)
A Comparison Of Some Models Of X-Ray Beam Hardening
Proceedings of SPIE (December 23 1976)
Optimizing Computed Tomography (CT) Scanner Geometry
Proceedings of SPIE (July 06 1979)

Back to Top