Paper
25 March 1976 X-Ray Automatic Exposure Timing And Control Circuitry
Edward S. Joseph, John E. Schneble
Author Affiliations +
Proceedings Volume 0070, Application of Optical Instrumentation in Medicine IV; (1976) https://doi.org/10.1117/12.954589
Event: Application of Optical Instrumentation in Medicine, IV, 1975, Atlanta, United States
Abstract
A block diagram representation of the circuitry is shown in Figures 1 and 2. Figure 1 shows the basic elements of an automatic exposure control with an ionization chamber as a detector. (Phototubes may also be used in place of ion chambers.) Figure 2 shows a typical automatic exposure control system with several ionization chamber detectors, and a photospot detector. It also shows an optional tomographic exposure control with current feedback and kVp servo compensation.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Edward S. Joseph and John E. Schneble "X-Ray Automatic Exposure Timing And Control Circuitry", Proc. SPIE 0070, Application of Optical Instrumentation in Medicine IV, (25 March 1976); https://doi.org/10.1117/12.954589
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal detection

Sensors

Ions

Medicine

Automatic exposure

Control systems

Tomography

Back to Top