25 July 1979 High Resolution Tv Fluoroscopy and Image Storing
Author Affiliations +
Proceedings Volume 0164, 4th European Electro-Optics Conf; (1979) https://doi.org/10.1117/12.965508
Event: Fourth European Electro-Optics Conference, 1978, Utrecht, Netherlands
Abstract
The investigation of the image-quality determining sections of the X-ray TV chain is an important matter for the optimization of the whole system. The modulation-transfer functions (MTF) of each component as the most important factor for the evaluation of the separate sections of the X-ray chain were measured. Special attention was given to gain exact measurements of MTF and other behavior of image pick-up tubes. In the second part of this paper image-quality determining factors and development tendencies of appropriate video stores for high resolution are discussed.
© (1979) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Haendle, J. Haendle, H. Sklebitz, H. Sklebitz, } "High Resolution Tv Fluoroscopy and Image Storing", Proc. SPIE 0164, 4th European Electro-Optics Conf, (25 July 1979); doi: 10.1117/12.965508; https://doi.org/10.1117/12.965508
PROCEEDINGS