Paper
7 October 1980 Light Levels In Ophthalmic Diagnostic Instruments
Francois C. Delori, Oleg Pomerantzeff, Martin A. Mainster
Author Affiliations +
Proceedings Volume 0229, Ocular Effects of Non-Ionizing Radiation; (1980) https://doi.org/10.1117/12.958799
Event: 1980 Technical Symposium East, 1980, Washington, D.C., United States
Abstract
Methods are presented for determining the retinal exposures produced by broad-band ophthalmic light sources and for comparing these results with safety standards developed for the safe use of lasers (ANSI, Z136.1). Measurements of exposures produced by fundus cameras and indirect ophthalmoscopes are compared with both Z136.1 maximum permissible exposure levels (MPE) and experimental retinal damage thresholds. It is found that while the retinal exposures used in fundus photography are below MPEs, the irradiances used in indirect ophthalmoscopy may actually equal or exceed current MPEs. Methods are suggested for reducing retinal irradiance while maintaining retinal image luminance, and emphasis is placed on the importance of both acquainting the users of ophthalmic devices with any potential retinal hazard and providing the users and designers of ophthalmic devices with realistic protection standards.
© (1980) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Francois C. Delori, Oleg Pomerantzeff, and Martin A. Mainster "Light Levels In Ophthalmic Diagnostic Instruments", Proc. SPIE 0229, Ocular Effects of Non-Ionizing Radiation, (7 October 1980); https://doi.org/10.1117/12.958799
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Light sources

Standards development

Photography

Retina

Diagnostics

Ophthalmoscopy

RELATED CONTENT


Back to Top