29 December 1982 Multiparameter Optimization Of Mammography
Author Affiliations +
Proceedings Volume 0347, Application of Optical Instrumentation in Medicine X; (1982) https://doi.org/10.1117/12.933811
Event: Application of Optical Instrumentation in Medicine X, 1982, New Orleans, United States
Abstract
Analytical expressions have been written for image quality in mammography. A multiparameter optimization has been conducted to find the mammography system requiring the lowest patient dose for a given image quality. The optimization is subject to constraints imposed by technology such as; power limits on tube focal spots, absorption efficiency related to detector resolution and others. The optimization permits system geometry, kVp, filtration, detector resolution, focal spot size and grid characteristics to vary simultaneously and self consistently subject to an exposure time constraint. Significant dose reductions compared to current mammography systems have been found without assuming radical technological advances.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
H. Jafroudi, H. Jafroudi, E. P. Muntz, E. P. Muntz, H. Bernstein, H. Bernstein, R. J. Jennings, R. J. Jennings, "Multiparameter Optimization Of Mammography", Proc. SPIE 0347, Application of Optical Instrumentation in Medicine X, (29 December 1982); doi: 10.1117/12.933811; https://doi.org/10.1117/12.933811