1 November 1982 Contrast Sensitivity Of Digital Fluorographic Systems
Author Affiliations +
Abstract
Several aspects of iodine sensitivity of digital fluorographic systems are discussed as applied to digital subtraction angiography. Expected iodine signal magnitudes are reviewed and shown to be in the range of 1-10 mg/cm2. An experimental example of spurious contrast in iodine sensitivity measurements is2 provided. Imaging of a 1 mm wide 0.25 mg/cm simulated iodinated vessel is demonstrated. Such iodine amounts are thought to be near the limit of the sensitivity of digital fluorographic systems. The concept of matched filtering is proposed for generating the equivalent of temporal subtraction images but using low intensity continuous x-ray exposures. The first experimental matched filter canine image is presented using a total exposure for the procedure about one-sixth of that used for pulsed exposure images of comparable quality.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. J. Riederer, S. J. Riederer, A. L. Hall, A. L. Hall, G. S. Keyes, G. S. Keyes, N. J. Pelc, N. J. Pelc, } "Contrast Sensitivity Of Digital Fluorographic Systems", Proc. SPIE 0372, Physics and Engineering in Medical Imaging, (1 November 1982); doi: 10.1117/12.934496; https://doi.org/10.1117/12.934496
PROCEEDINGS
8 PAGES


SHARE
RELATED CONTENT

Video signal processing using vector median
Proceedings of SPIE (August 31 1990)
Digital Subtraction Arteriography: Now And The Future
Proceedings of SPIE (December 28 1982)
Doppler-guided retrograde catheterization system
Proceedings of SPIE (April 30 1991)
Iodine Sensitivity Of Digital Imaging Systems
Proceedings of SPIE (November 03 1981)
Intravenous Contrast Angiography
Proceedings of SPIE (June 18 1979)

Back to Top