Translator Disclaimer
1 November 1982 The Imaging Performance Of A Multiwire Proportional Chamber Positron Camera(+)
Author Affiliations +
Proceedings Volume 0375, Medical Imaging and Image Interpretation; (1982) https://doi.org/10.1117/12.934606
Event: 1st International Symposium on Medical Imaging and Image Interpretation, 1982, Berlin, Germany
Abstract
A new design - fully three dimensional -Positron Camera is presented, made of six MultiWire Proportional Chamber modules arranged to form the lateral surface of a hexagonal prism. A true coincidence rate of 56000 c/s is expected with an equal accidental rate for a 400 iCi activity uniformly distributed in a --3k water phantom. A detailed Monte Carlo program has been used to investigate the dependence of the spatial resolution on the geometrical and physical parameters. A spatial resolution of 4.8 mm FWHM has been obtained for a 18F point-like source in a 10 cm radius water phantom. The main properties of the limited angle reconstruction algorithms are described in relation to the proposed detector geometry.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Victor Perez-Mendez, Alberto Del Guerra, Walter R. Nelson, and Kwok C. Tam "The Imaging Performance Of A Multiwire Proportional Chamber Positron Camera(+)", Proc. SPIE 0375, Medical Imaging and Image Interpretation, (1 November 1982); https://doi.org/10.1117/12.934606