13 December 1983 Optimal Dose Utilization With Variable X-Ray Intensity In Digital Radiography
Dennis L. Parker, Paul D. Clayton, Lawrence R. Tarbox, Patrick L. VonBehren
Author Affiliations +
Proceedings Volume 0419, Application of Optical Instrumentation in Medicine XI; (1983) https://doi.org/10.1117/12.936012
Event: Application of Optical Instrumentation in Medicine XI, 1983, Atlanta, United States
Abstract
For the case of radiographic contrast media flowing through an otherwise stationary object, a theoretical analysis shows that image enhancement with optimal dose utilization can be achieved by varying the x-ray intensity during the sequence of exposures. The need for variable intensity derives from the conflicting requirements of continuous visualization of the time course of the contrast media and the fact that dose utilization in a difference image is maximized when all the x-ray photons are divided between minimum and maximum opacification. A set of optimal weights for combining multiple images are derived as a function of the x-ray intensity in each image and the effects of photon noise, camera (video system) noise, and digital truncation noise. It is shown theoretically that controlled variation of x-ray intensity may improve the signal to noise to dose ratio by approximately a factor of 2 when compared to conventional matched filtering. If the x-ray intensity remains constant for all images, the optimal filter weights reduce to those of the expected matched filter. A simple implementation of optimal filtering using only two intensities is described mathematically and demonstrated experimentally.
© (1983) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Dennis L. Parker, Paul D. Clayton, Lawrence R. Tarbox, and Patrick L. VonBehren "Optimal Dose Utilization With Variable X-Ray Intensity In Digital Radiography", Proc. SPIE 0419, Application of Optical Instrumentation in Medicine XI, (13 December 1983); https://doi.org/10.1117/12.936012
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

X-ray imaging

Image filtering

Digital filtering

Image processing

Cameras

Signal to noise ratio

RELATED CONTENT

Tomographic Filters For Digital Radiography
Proceedings of SPIE (November 04 1981)
Enhancement of x-ray fluoroscopy images
Proceedings of SPIE (September 14 1993)
Toward real-time clamping load measurement using DSPI
Proceedings of SPIE (September 13 2012)
Energy-resolved soft x-ray framing camera
Proceedings of SPIE (August 01 2003)

Back to Top