20 December 1985 Thermal CalibratiOn of a Low Temperature Invar Sample Cell for FTIR Spectroscopy
Author Affiliations +
Proceedings Volume 0553, Fourier and Computerized Infrared Spectroscopy; (1985) https://doi.org/10.1117/12.970924
Event: 1985 International Conference on Fourier and Computerized Infrared Spectroscopy, 1985, Ottawa, Canada
Abstract
A sample cell, capable of supporting standard 13 mm diameter pressed pellets, and suitable for mounting in a commercial cryostat, has been constructed of Invar, and fitted with two thermocouple sensors for operating in the range 4.2 < T < 300 K. The thermal lag between the temperature of the experimental sample and the sensor reading has been determined by monitoring four phase changes in covalent solids which can be used as "infrared thermometers". In the temperature range 77 < T < 300 K this thermal lag is less than 0.5 degrees.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
L. M. Casson, L. M. Casson, R. H. Herber, R. H. Herber, } "Thermal CalibratiOn of a Low Temperature Invar Sample Cell for FTIR Spectroscopy", Proc. SPIE 0553, Fourier and Computerized Infrared Spectroscopy, (20 December 1985); doi: 10.1117/12.970924; https://doi.org/10.1117/12.970924