19 September 1985 Utilization Of The Monte Carlo Technique To Assess Precision In The Measurement Of Coronary Arteries
Author Affiliations +
Abstract
Using standard angiographic equipment involving an image intensifier-video camera combination, subpixel precision can be achieved in the estimation of blood vessel diameter using 512x512x8 digital image resolution. Analyses were carried out with blood vessel phantoms ranging from 1.27 to 5.46 mm in diameter. The relationship between actual and measured diameters was found linear between 3 and 5.46 mm; below 3mm it became non-linear. The studies were performed under ideal conditions with a water block as scattering medium, without motion and without subtraction on uniform background.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tamas Sandor, Tamas Sandor, J.Richard Spears, J.Richard Spears, Lawrence M. Boxt, Lawrence M. Boxt, David C. Levin, David C. Levin, William B. Hanlon, William B. Hanlon, Donald P. Harrington, Donald P. Harrington, Andrew P . Selwyn, Andrew P . Selwyn, Sven Paulin, Sven Paulin, } "Utilization Of The Monte Carlo Technique To Assess Precision In The Measurement Of Coronary Arteries", Proc. SPIE 0555, Medical Imaging and Instrumentation '85, (19 September 1985); doi: 10.1117/12.949489; https://doi.org/10.1117/12.949489
PROCEEDINGS
6 PAGES


SHARE