12 August 1986 X-ray Microtomography with Synchrotron Radiation
Author Affiliations +
Abstract
High resolution x-ray microtomography has been established at the synchrotron radiation source at Daresbury Laboratory, U.K. A high precision mechanical stage with two translational and one rotational degree of freedom has been constructed, and fully characterised by mechanical metrology; overall precision is better than 0.5 μm. Collimation of the incident synchrotron radiation beam through a 4 pm aperture has been obtained, giving 2 μm overall spatial resolution. Results have been obtained using monochromatic radiation at both 1 angstrom and 0.37 angstrom wavelengths, on human bone fragments and on sintered alumina ceramics.
© (1986) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
D. K. Bowen, D. K. Bowen, J. S. Elliott, J. S. Elliott, S. R. Stock, S. R. Stock, S. D. Dover, S. D. Dover, } "X-ray Microtomography with Synchrotron Radiation", Proc. SPIE 0691, X-Ray Imaging II, (12 August 1986); doi: 10.1117/12.936625; https://doi.org/10.1117/12.936625