1 January 1986 Spectral Efficiency and Resolution of Si(Li)-Detectors for Photon Energies between 0.3 keV and 5 keV
Author Affiliations +
Proceedings Volume 0733, Soft X-Ray Optics and Technology; (1986); doi: 10.1117/12.964950
Event: Soft X-Ray Optics and Technology, 1987, Berlin, Germany
Abstract
The spectral efficiencies of energy dispersive Si(Li) photon counters have been measured using the storage ring BESSY as a radiometric standard source of extremely low photon flux of the order of 1 photon/(s eV). The detectors were irradiated with white synchrotron radia-tion when the storage ring was operated with only about 5 electrons stored. For energy calibration and measurement of the energy resolution X-ray emission lines excited by a 55-Fe source were used. Towards lower photon energies the efficiency is drastically decreased by the building-up of an ice-layer on the permanently cooled detector. By this surface contamination also the energy resolution of the detector is affected. The performance of detector can be recovered by a warming-up procedure.
© (1986) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
F. Riehle, E. Tegeler, B. Wende, "Spectral Efficiency and Resolution of Si(Li)-Detectors for Photon Energies between 0.3 keV and 5 keV", Proc. SPIE 0733, Soft X-Ray Optics and Technology, (1 January 1986); doi: 10.1117/12.964950; https://doi.org/10.1117/12.964950