1 January 1987 Theoretical Considerations Of A New Electronically Collimated Gamma Camera Utilizing Gas Scintillation
Author Affiliations +
Abstract
A new electronically collimated gamma camera utilizing a gas scintillation position-sensitive detector and a multiwire proportional chamber is proposed and its imaging characteristics are discussed in this paper. The scheme preserves all the advantages of an electronically collimated system (ECS) i.e. high sensitivity and simultaneous multiple views of the object over the conventional NaI gamma camera. Compared with the Ge based ECS, this scheme would have higher spatial resolution and avoid the construction difficulties of a large area Ge detector.
© (1987) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ichiro Fujieda and V. Perez-Mendez "Theoretical Considerations Of A New Electronically Collimated Gamma Camera Utilizing Gas Scintillation", Proc. SPIE 0767, Medical Imaging, (1 January 1987); https://doi.org/10.1117/12.966984