4 February 1988 A Soft X-Ray Streak Camera Using A Microchannel Plate Photocathode
Author Affiliations +
Abstract
We present data characterizing an x-ray streak camera which uses a microchannel plate as a photocathode. The microchannel plate provides good sensitivity to soft x-ray photons and adjustable gain in addition to one dimensional spatial imaging. The streak camera was used in conjunction with a high sensitivity CCD camera and a computer data acquisition system. The resulting detection system has single photon sensitivity, a time resolution of better than 200 psec, and a dynamic range exceeding 103.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
H. C. Kapteyn, H. C. Kapteyn, W. W. Craig, W. W. Craig, G. D. Power, G. D. Power, J. Schachter, J. Schachter, R. W. Falcone, R. W. Falcone, } "A Soft X-Ray Streak Camera Using A Microchannel Plate Photocathode", Proc. SPIE 0832, High Speed Photography, Videography, and Photonics V, (4 February 1988); doi: 10.1117/12.942254; https://doi.org/10.1117/12.942254