17 February 2017 Optical coherence tomography with pre-calculated reference spectra
Author Affiliations +
Abstract
The resample of spectra which is essential for high-precision spectral-domain OCT data processing is sophisticated, and its precision is dependent on the method and equipment. In this paper, we proposed an OCT without inverse FFT. A series of reference spectra corresponding to different optical path length difference was used to convolve with spectra gotten by OCT to acquire time-domain tomography instead of inverse FFT, thus eliminating the resample of spectra. The reference spectra were calculated before imaging and corrected with correction spectrum from sample arm to compensate the influence of sample arm. Experiment was done with a mirror as sample and validated our setup.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yi Wang, Yi Wang, Xiaodong Chen, Xiaodong Chen, Haitao Liang, Haitao Liang, Daoyin Yu, Daoyin Yu, } "Optical coherence tomography with pre-calculated reference spectra", Proc. SPIE 10053, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXI, 1005330 (17 February 2017); doi: 10.1117/12.2242359; https://doi.org/10.1117/12.2242359