21 February 2017 Measurement of tissue viscoelasticity with ultrasound
Author Affiliations +
Abstract
Tissue properties such as elasticity and viscosity have been shown to be related to such tissue conditions as contraction, edema, fibrosis, and fat content among others. Magnetic Resonance Elastography has shown outstanding ability to measure the elasticity and in some cases the viscosity of tissues, especially in the liver, providing the ability to stage fibrotic liver disease similarly to biopsy. We discuss ultrasound methods of measuring elasticity and viscosity in tissues. Many of these methods are becoming widely available in the extant ultrasound machines distributed throughout the world. Some of the methods to be discussed are in the developmental stage. The advantages of the ultrasound methods are that the imaging instruments are widely available and that many of the viscoelastic measurements can be made during a short addition to the normal ultrasound examination time. In addition, the measurements can be made by ultrasound repetitively and quickly allowing evaluation of dynamic physiologic function in circumstances such as muscle contraction or artery relaxation. Measurement of viscoelastic tissue mechanical properties will become a consistent part of clinical ultrasound examinations in our opinion.
Conference Presentation

1.

Introduction

For numerous decades, the physical material properties of soft tissue have been an important topic of investigation. Several of the imaging techniques image fundamental attributes such as differences of mass density in x-ray computed tomography; proton density among other molecular properties in magnetic resonance imaging; and backscattering characteristics based on differences in acoustic impedance in ultrasonic imaging. The field of elastic or elasticity imaging in medical imaging modality is based on the hypothesis that alterations in tissue state are mirrored by alterations in the mechanical viscoelastic properties of soft tissues(1). Measurements of the mechanical properties of soft tissues can be done several ways, however the methods generally consist of two types – the application of a force or stress from the outside of the body and then imaging the resulting deformation (4), or applying force inside the body while measuring and imaging the deformation. Measurement of the deformation can be done using ultrasound or magnetic resonance imaging (3,7).

2.

Viscoelastic parameter estimation

Quantitative elasticity imaging requires determination of the elastic or viscoelastic modulus of the tissue (2). Here, we derive some relationships between what can be measured using magnetic resonance imaging or ultrasound and fundamental properties, such as the Young’s Modulus, which is the ratio of unilateral stress to strain in a sample. The Young’s modulus is defined for compressional stress and the shear modulus is defined for shear stress, which is also the ratio of shear stress to shear strain. Young’s modulus E and shear modulus G are related by the Possion ratio v where;

00006_PSISDG10067_1006705_page_1_1.jpg

In the following derivations, we confine our studies to soft tissues, which are water saturated, and therefore, extremely incompressible giving a Possion ratio of about 0.5. This produces a Young’s modulus equal to 3 times the shear modulus. The shear modulus can be considered complex e.g.,

00006_PSISDG10067_1006705_page_1_2.jpg

where ω = 2πf is the angular frequency, 𝘎1 storage modulus, 𝘎2 is the loss modulus. Therefore, the complex shear modulus can also be written as

00006_PSISDG10067_1006705_page_2_1.jpg

where ρ is the mass density of the medium. In soft tissues, this is assumed to be ρ = 1000 kg/m2 and k is the complex wave number. Shear wave velocity and shear wave attenuation can be written as these two terms:

00006_PSISDG10067_1006705_page_2_2.jpg
00006_PSISDG10067_1006705_page_2_3.jpg

Viscoelastic materials can be modeled using a rheological approach. A widely used model in elasticity imaging is the Kelvin-Voigt model, which includes an elastic spring and a viscous dashpot in parallel and provides shear elasticity µ1 and viscosity µ2.

Shear wave velocity and shear wave attenuation can be computed giving the next 2 equations;

00006_PSISDG10067_1006705_page_2_4.jpg
00006_PSISDG10067_1006705_page_2_5.jpg

The implications of these two equations are that shear wave velocities and attenuation are functions of frequency and the elastic and viscous terms of the medium. The above two equations can be solved for µ1 and µ2 if we can measure the speed cs and attenuation αs over a range of frequencies.

3.

The shear wave approach

Sarvazyan et. al. suggested a method called shearwave elasticity imaging, SWEI (1). In this method radiation force produced by a focused ultrasound creates shearwaves in the tissue. This shearwave propagates out from the focal center. The speed of these shearwaves can be related through the equation 00006_PSISDG10067_1006705_page_2_6.jpg if there is no attenuation.

There are many methods of creating shearwaves including those by SSI (8), in which short high intensity ultrasound tone bursts are focused along a line of points in the tissue and producing shearwaves. Imaging the shearwave propagation requires high speed ultrasound to measure particle motion because the shear wave propagates in tissue at up to 5 meters per second. The most advanced modern ultrasound systems can obtain up to 10,000 frames per second and so these shearwaves can be measured rather easily. Shearwave velocity dispersion in various tissues have been reported previously in such tissues and organs as liver, kidney, brain, prostate, skeletal muscle, breast, cornea, and blood clots(10,13,14,19, 20, 22, 23).

3.1.

SDUV

A new method of measuring shearwaves is to use shearwave dispersion ultrasound velocimetry, SDUV (12, 19). The speed of the shearwave is calculated by the following equation:

00006_PSISDG10067_1006705_page_2_7.jpg

where Δ∅ and Δr are phase difference and distance between two observation point along the shearwave path. Linear regression can be used to calculate a least square estimate of the ratio for each frequency. This measurement of course is made over a range of frequencies because the speed of the shearwave varies if the velocity is dispersive, that is, the imaginary term K2(ω) is not 0.

We have developed a new method of inducing many shearwaves in tissue by pushing multiple areas simultaneously, thus, shearwaves propagate in all directions at the same time. Directional filtering is used to produce sequential images of shearwaves propagating in selected directions. The speed of propagation of these shearwaves is used to calculate the material properties of the tissue. This method is called CUSE for comb-beam ultrasound shear elastography. There have been several applications of comb-beam elastography.

The following is an example clinical trial in which CUSE has been used to diagnose lymph node metastasis in patients with breast cancer. In a set of 68 patients whose suspicious lymph nodes were detected with ultrasound, half were found to be negative and half were positive by biopsy. CUSE was used to determine the speed/stiffness of the nodes as shown in Figures 1 and 2. Table 1 shows the means and standard deviations of the set of stiffness measurements in the positive and negative lymph nodes. Setting the sensitivity of the CUSE shear wave method at 100% gives a specificity of 84%. This method would have saved about 26 biopsies if it were used on the original set of 68 patients.

Figure 1.

B-scan, left, and tissue stiffness scan, right, of breast lesions biopsied as reactive lymph node. The speed of shear wave ranges from 3.25 to 0.86 m/s.

00006_PSISDG10067_1006705_page_3_1.jpg

Figure 2.

B-scan of lymph gland, left, and stiffness scan, right, of biopsied proven adenocarcinoma. Speed of shear wave ranges from 8.6 to 2.3 m/s.

00006_PSISDG10067_1006705_page_4_1.jpg

Table 1

shows the effect of malignancy on stiffness for the 68 patients.

4.

Conclusion

Measurements of tissue mechanical properties using ultrasound can be used to diagnose metastatic lymph nodes in patients with breast cancer. After detection of suspicious lymph nodes with ultrasound imaging, quantitative measurements of stiffness using the CUSE method can diagnose biopsy proven cancerous/benign nodes with 100% sensitivity and 84% specificity possibly saving many biopsies if used in practice. The wide availability of ultrasound methods and their ubiquitous use in breast ultrasound examinations provides a universal basis for the application of this method in the diagnosis of malignant metastasis in lymph nodes of these patients.

Acknowledgments

The project described was supported in part by NIH grant R01 CA148994. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NCI or NIH. Mayo Clinic and the presenter have financial interest from the licensing of a product or technology related to this presentation.

References

[1] 

Sarvazyan, A., Hall, T.J., Urban, M.W., Fatemi, M., Aglyamov, S.R., Garra, B.S., “Elasticity imaging – an emerging branch of medical imaging. An overview,” Curr Med Imaging Rev., 7 (4), 255 –282 (2011). https://doi.org/10.2174/157340511798038684 Google Scholar

[2] 

Chen, S., Urban, M.W., Pislaru, C., Kinnick, R., Zheng, Y., Yao, A., Greenleaf, J.F., “Shearwave dispersion ultrasound vibrometry (SDUV) for measuring tissue elasticity and viscosity,” IEEE Trans Ultrason Ferroelectr Freq Control, 56 55 –62 (2009). https://doi.org/10.1109/TUFFC.2009.1005 Google Scholar

[3] 

Muthupillai, R., Lomas, D.J., Rossman, P.J., Greenleaf, J.F., Manduca, A., Ehman, R.L., “Magnetic resonance elastography by direct visualization of propagating acoustic strain waves,” Science, 269 1854 –1857 (1995). https://doi.org/10.1126/science.7569924 Google Scholar

[4] 

Lerner, R.M., Huang, S.R., Parker, K.J., ““Sonoelasticity” images derived from ultrasound signals in mechanically vibrated tissues,” Ultrasound Med Biol, 16 231 –9 (1990). https://doi.org/10.1016/0301-5629(90)90002-T Google Scholar

[5] 

Sandrin, L., Catheline, S., Tanter, M., Hennequin, X., Fink, M., “Time-resolved pulsed elastography with ultrafast ultrasonic imagin,” Ultrason Imaging, 21 259 –72 (1999). https://doi.org/10.1177/016173469902100402 Google Scholar

[6] 

Bercoff, J., Tanter, M., Fink, M., “Supersonic shear imaging: a new technique for soft tissue elasticity mapping,” IEEE Trans Ultrason Ferroelectr Freq Control, 51 396 –409 (2004). https://doi.org/10.1109/TUFFC.2004.1295425 Google Scholar

[7] 

Sinkus, R., Tanter, M., Xydeas, T., Catheline, S., Bercoff, J., Fink, M., “Viscoelastic shear properties of in vivo breast lesions measured by MR elastography,” Magn Reson Imaging, 23 159 –65 (2005). https://doi.org/10.1016/j.mri.2004.11.060 Google Scholar

[8] 

Tanter, M., Bercoff, J., Athanasiou, A., Deffieux, T., Gennisson, J. L., Montaldo, G., Muller, M., Tardivon, A., Fink, M., “Quantitative assessment of breast lesion viscoelasticity: Initial clinical results using supersonic shear imaging,” Ultrasound Med Biol, 34 1373 –1386 (2008). https://doi.org/10.1016/j.ultrasmedbio.2008.02.002 Google Scholar

[9] 

Deffieux, T., Montaldo, G., Tanter, M., Fink, M., “Shear wave spectroscopy for in vivo quantification of human soft tissues visco-elasticity,” IEEE Trans Med Imaging, 28 313 –322 (2009). https://doi.org/10.1109/TMI.2008.925077 Google Scholar

[10] 

Muller, M., Gennisson, J.L., Deffieux, T., Tanter, M., Fink, M., “Quantitative viscoelasticity mapping of human liver using supersonic shear imaging: preliminary in vivo feasibility study,” Ultrasound Med Biol, 35 219 –229 (2009). https://doi.org/10.1016/j.ultrasmedbio.2008.08.018 Google Scholar

[11] 

Tanter, M., Touboul, D., Gennisson, J.L., Bercoff, J., Fink, M., “High-resolution quantitative imaging of cornea elasticity using supersonic shear imaging,” IEEE Trans Med Imaging, 28 1881 –1893 (2009). https://doi.org/10.1109/TMI.2009.2021471 Google Scholar

[12] 

Chen, S., Fatemi, M., Greenleaf, J.F., “Quantifying elasticity and viscosity from measurement of shear wave speed dispersion,” J Acoust Soc Am, 115 2781 –5 (2004). https://doi.org/10.1121/1.1739480 Google Scholar

[13] 

Zhang, X., Greenleaf, J.F., “The stiffening of arteries by the tissue-mimicking gelatin,” IEEE Trans Ultrason Ferroelectr Freq Control, 53 1534 –9 (2006). https://doi.org/10.1109/TUFFC.2006.1665111 Google Scholar

[14] 

Zhang, X., Greenleaf, J.F., “Measurement of wave velocity in arterial walls with ultrasound transducers,” Ultrasound Med Biol, 32 1655 –60 (2006). https://doi.org/10.1016/j.ultrasmedbio.2006.04.004 Google Scholar

[15] 

Zhang, X.M., Greenleaf, J.F., “Estimation of tissue’s elasticity with surface wave speed,” J Acoust Soc Am, 122 2522 –2525 (2007). https://doi.org/10.1121/1.2785045 Google Scholar

[16] 

Urban, M.W., Greenleaf, J.F., “Harmonic pulsed excitation and motion detection of a vibrating reflective target,” J Acoust Soc Am, 123 519 –533 (2008). https://doi.org/10.1121/1.2805666 Google Scholar

[17] 

Zhang, X., Kinnick, R.R., Fatemi, M., Greenleaf, J.F., “Noninvasive method for estimation of complex elastic modulus of arterial vessels,” IEEE Trans Ultrason Ferroelectr Freq Control, 52 642 –652 (2005). https://doi.org/10.1109/TUFFC.2005.1428047 Google Scholar

[18] 

Zhang, X., Greenleaf, J.F., “Noninvasive generation and measurement of propagating waves in arterial walls,” J Acoust Soc Am, 119 1238 –43 (2006). https://doi.org/10.1121/1.2159294 Google Scholar

[19] 

Mitri, F.G., Urban, M.W., Fatemi, M., Greenleaf, J.F., “Shearwave Dispersion Ultrasonic Vibrometry for measuring prostate shear stiffness and viscosity – An in vitro pilot study,” IEEE Trans Biomed Eng., 58 235 –242 (2011). https://doi.org/10.1109/TBME.2010.2053928 Google Scholar

[20] 

Nenadic, I., Urban, M.W., Greenleaf, J.F., “Study of Lamb wave dispersion in porcine myocardium,” J Acoust Soc Am, 124 2445 –2445 (2008). https://doi.org/10.1121/1.4782583 Google Scholar

[21] 

Nenadic, I.; Urban, M.W.; Greenleaf, JF., “Ex Vivo measurements of myocardial viscoelasticity using Shearwave Dispersion Ultrasound Vibrometry (SDUV),” Engineering in Medicine and Biology Society, 2895 –2898 (2009). Google Scholar

[22] 

Pislaru, C.; Urban, M.W.; Nenadic, I.; Greenleaf, J.F., “Shearwave dispersion ultrasound vibrometry applied to in vivo myocardium,” Engineering in Medicine and Biology Society, 2891 –2894 (2009). Google Scholar

[23] 

Athanasiou, A., Tardivon, A., Tanter, M.l., Sigal-Zafrani, B., Bercoff, J., Deffieux, T., Gennisson, J-L, Fink, M., Neuenschwander S., “Breast lesions: quantitative elastography with supersonic shear imaging: preliminary results,” Radiology, 256 297 –303 (2010). https://doi.org/10.1148/radiol.10090385 Google Scholar
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. F. Greenleaf and A. Alizad M.D. "Measurement of tissue viscoelasticity with ultrasound", Proc. SPIE 10067, Optical Elastography and Tissue Biomechanics IV, 1006705 (21 February 2017); https://doi.org/10.1117/12.2256842