You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
24 February 2017Image enhancement in positron emission mammography
Purpose: To evaluate an efficient iterative deconvolution method (RSEMD) for improving the quantitative accuracy
of previously reconstructed breast images by commercial positron emission mammography (PEM) scanner.
Materials and Methods: The RSEMD method was tested on breast phantom data and clinical PEM imaging data.
Data acquisition was performed on a commercial Naviscan Flex Solo II PEM camera. This method was applied to
patient breast images previously reconstructed with Naviscan software (MLEM) to determine improvements in
resolution, signal to noise ratio (SNR) and contrast to noise ratio (CNR.)
Results: In all of the patients’ breast studies the post-processed images proved to have higher resolution and lower
noise as compared with images reconstructed by conventional methods. In general, the values of SNR reached a
plateau at around 6 iterations with an improvement factor of about 2 for post-processed Flex Solo II PEM images.
Improvements in image resolution after the application of RSEMD have also been demonstrated.
Conclusions: A rapidly converging, iterative deconvolution algorithm with a novel resolution subsets-based
approach RSEMD that operates on patient DICOM images has been used for quantitative improvement in breast
imaging. The RSEMD method can be applied to clinical PEM images to improve image quality to diagnostically
acceptable levels and will be crucial in order to facilitate diagnosis of tumor progression at the earliest stages. The
RSEMD method can be considered as an extended Richardson-Lucy algorithm with multiple resolution levels
(resolution subsets).
The alert did not successfully save. Please try again later.
Nikolai V. Slavine, Stephen Seiler, Roderick W. McColl, Robert E. Lenkinski, "Image enhancement in positron emission mammography," Proc. SPIE 10133, Medical Imaging 2017: Image Processing, 101331J (24 February 2017); https://doi.org/10.1117/12.2248758