28 February 2017 Calibration on the detection efficiency of the Si-APD and InGaAs-APD single-photon detectors by correlated photon pairs
Author Affiliations +
Proceedings Volume 10256, Second International Conference on Photonics and Optical Engineering; 1025622 (2017) https://doi.org/10.1117/12.2257323
Event: Second International Conference on Photonics and Optical Engineering, 2016, Xi'an, China
Abstract
We demonstrated calibration on the detection efficiency of Si-avalanche photodiode (APD) and InGaAs-APD singlephoton detectors by correlated photon pairs at 780 nm and 1550 nm, respectively. The correlated photons were generated by spontaneous frequency down-conversion in a periodically poled potassium titanyl phosphate crystal (PPKTP) pumped by a pulsed fiber laser. The uncertainty of ~10-4 on detection efficiency was obtained for both single-photon detectors.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xueshun Shi, Xueshun Shi, Kun Zhao, Kun Zhao, Changming Liu, Changming Liu, Haidong Chen, Haidong Chen, Kunfeng Chen, Kunfeng Chen, Haiyong Gan, Haiyong Gan, } "Calibration on the detection efficiency of the Si-APD and InGaAs-APD single-photon detectors by correlated photon pairs", Proc. SPIE 10256, Second International Conference on Photonics and Optical Engineering, 1025622 (28 February 2017); doi: 10.1117/12.2257323; https://doi.org/10.1117/12.2257323