10 January 1989 Beam delivery systems
Author Affiliations +
Proceedings Volume 10305, Dosimetry of Laser Radiation in Medicine and Biology; 103050F (1989) https://doi.org/10.1117/12.2283596
Event: Institutes for Advanced Optical Technologies, 1988, Berlin, Germany
Abstract
The delivery device attached to any medical laser system is the very important, if somewhat neglec- ted, "Cinderella" of the system. It is the part that the surgeon or operator interarts with and yet de- spite occupying this key position there are few, if any, objective standards to which the delivery de- vices have to conform. There is a requirement, for example, that if the power output from a CO2 laser varies outside the limits set at + 20 % a system should at least indicate a fault. Despite providing some indication of the output power calibrated in such a way as to represent the output power of the end of the articulating arm there is no requirement on continuous monitoring at the end of the arm and therefore, should a mirror fail through being contaminated with smoke that variation would not be recorded nor any action taken unless the drop in power was noted by the surgeon. I am not necessarily argueing in favour of distal power monitoring but simply make the point that failures of the delivery system have not yet been incorporated into formal regulations.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. J. Colles, M. J. Colles, } "Beam delivery systems", Proc. SPIE 10305, Dosimetry of Laser Radiation in Medicine and Biology, 103050F (10 January 1989); doi: 10.1117/12.2283596; https://doi.org/10.1117/12.2283596