7 June 1989 Measurements Of The Temporal And Spatial Resolutions Of X-Ray Streak Camera
Author Affiliations +
Proceedings Volume 1032, 18th Intl Congress on High Speed Photography and Photonics; (1989) https://doi.org/10.1117/12.969124
Event: 18th International Congress on High Speed Photography and Photonics, 1988, Xi'an, Shaanxi, China
Abstract
The methods and the resits of the measurements of the temporal and spatial resolutions of our X-ray streak camera are described. The temporal resolution measurement was performed by the method of laser produced plasma. Multilayer filters were used to remove the softer and longer X-ray pulses. X-ray pulse down to 33 ps in duration has been measured. An ARR CPM Nd:YAG oscillator was used to get even shorter laser pulses and preliminary result was obtained. The spatial resolution was measured by using grid form photocathode. The static spatial resolution of our camera up to 20 1p/mm and dynamic spatial resolution of 8 1p/mm have been realized.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Qiushi Gao, Qiushi Gao, Zhiyuan Lei, Zhiyuan Lei, Zhongxian Song, Zhongxian Song, } "Measurements Of The Temporal And Spatial Resolutions Of X-Ray Streak Camera", Proc. SPIE 1032, 18th Intl Congress on High Speed Photography and Photonics, (7 June 1989); doi: 10.1117/12.969124; https://doi.org/10.1117/12.969124
PROCEEDINGS
5 PAGES


SHARE
RELATED CONTENT


Back to Top