23 April 2017 Calibration and uncertainty of a fibre optic measurement system for Fresnel reflectometer sensors
Author Affiliations +
Proceedings Volume 10323, 25th International Conference on Optical Fiber Sensors; 103235O (2017) https://doi.org/10.1117/12.2260877
Event: 25th International Conference on Optical Fiber Sensors, 2017, Jeju, Korea, Republic of
Abstract
A four-channel fibre optic measurement system for fibre Bragg grating (FBG) sensors was used to measure the refractive index of fluids with Fresnel reflectometer sensors prepared from SMF28e singlemode fibres. The fibre optic measurement system was calibrated with five high-purity organic fluid samples in the refractive index range between 1.3 and 1.6. The light source had a wavelength of 1550 nm and the fluid samples were held at 24 °C. In total, 400 refractive index measurements were performed. The calibration of the fibre optic measurement system was performed with a linear correction function. The combined uncertainty of the refractive index measurements with the fibre optic measurement system was found to be in the order of 10-4.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jonathan Oelhafen, Jonathan Oelhafen, Tobias Mayr, Tobias Mayr, Konstantinos Moutzouris, Konstantinos Moutzouris, Johannes Roths, Johannes Roths, Klaus Drechsler, Klaus Drechsler, } "Calibration and uncertainty of a fibre optic measurement system for Fresnel reflectometer sensors", Proc. SPIE 10323, 25th International Conference on Optical Fiber Sensors, 103235O (23 April 2017); doi: 10.1117/12.2260877; https://doi.org/10.1117/12.2260877
PROCEEDINGS