29 August 2017 An automatic holographic adaptive phoropter
Author Affiliations +
Abstract
Phoropters are the most common instrument used to detect refractive errors. During a refractive exam, lenses are flipped in front of the patient who looks at the eye chart and tries to read the symbols. The procedure is fully dependent on the cooperation of the patient to read the eye chart, provides only a subjective measurement of visual acuity, and can at best provide a rough estimate of the patient’s vision. Phoropters are difficult to use for mass screenings requiring a skilled examiner, and it is hard to screen young children and the elderly etc. We have developed a simplified, lightweight automatic phoropter that can measure the optical error of the eye objectively without requiring the patient’s input. The automatic holographic adaptive phoropter is based on a Shack-Hartmann wave front sensor and three computercontrolled fluidic lenses. The fluidic lens system is designed to be able to provide power and astigmatic corrections over a large range of corrections without the need for verbal feedback from the patient in less than 20 seconds.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Babak Amirsolaimani, Babak Amirsolaimani, N. Peyghambarian, N. Peyghambarian, Jim Schwiegerling, Jim Schwiegerling, Arkady Bablumyan, Arkady Bablumyan, Nickolaos Savidis, Nickolaos Savidis, Gholam Peyman, Gholam Peyman, } "An automatic holographic adaptive phoropter", Proc. SPIE 10352, Biosensing and Nanomedicine X, 1035208 (29 August 2017); doi: 10.1117/12.2276807; https://doi.org/10.1117/12.2276807