7 September 2017 A framework for optimizing micro-CT in dual-modality micro-CT/XFCT small-animal imaging system
Author Affiliations +
Abstract
Dual-modality Computed Tomography (CT)/X-ray Fluorescence Computed Tomography (XFCT) can be a valuable tool for imaging and quantifying the organ and tissue distribution of small concentrations of high atomic number materials in small-animal system. In this work, the framework for optimizing the micro-CT imaging system component of the dual-modality system is described, either when the micro-CT images are concurrently acquired with XFCT and using the x-ray spectral conditions for XFCT, or when the micro-CT images are acquired sequentially and independently of XFCT. This framework utilizes the cascaded systems analysis for task-specific determination of the detectability index using numerical observer models at a given radiation dose, where the radiation dose is determined using Monte Carlo simulations.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Srinivasan Vedantham, Suman Shrestha, Andrew Karellas, Sang Hyun Cho, "A framework for optimizing micro-CT in dual-modality micro-CT/XFCT small-animal imaging system", Proc. SPIE 10393, Radiation Detectors in Medicine, Industry, and National Security XVIII, 103930R (7 September 2017); doi: 10.1117/12.2279351; https://doi.org/10.1117/12.2279351