7 September 2017 A framework for optimizing micro-CT in dual-modality micro-CT/XFCT small-animal imaging system
Author Affiliations +
Abstract
Dual-modality Computed Tomography (CT)/X-ray Fluorescence Computed Tomography (XFCT) can be a valuable tool for imaging and quantifying the organ and tissue distribution of small concentrations of high atomic number materials in small-animal system. In this work, the framework for optimizing the micro-CT imaging system component of the dual-modality system is described, either when the micro-CT images are concurrently acquired with XFCT and using the x-ray spectral conditions for XFCT, or when the micro-CT images are acquired sequentially and independently of XFCT. This framework utilizes the cascaded systems analysis for task-specific determination of the detectability index using numerical observer models at a given radiation dose, where the radiation dose is determined using Monte Carlo simulations.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Srinivasan Vedantham, Srinivasan Vedantham, Suman Shrestha, Suman Shrestha, Andrew Karellas, Andrew Karellas, Sang Hyun Cho, Sang Hyun Cho, } "A framework for optimizing micro-CT in dual-modality micro-CT/XFCT small-animal imaging system", Proc. SPIE 10393, Radiation Detectors in Medicine, Industry, and National Security XVIII, 103930R (7 September 2017); doi: 10.1117/12.2279351; https://doi.org/10.1117/12.2279351