6 September 2017 Method and algorithm for efficient calibration of compressive hyperspectral imaging system based on a liquid crystal retarder
Author Affiliations +
Abstract
Recently we presented a Compressive Sensing Miniature Ultra-spectral Imaging System (CS-MUSI)1 . This system consists of a single Liquid Crystal (LC) phase retarder as a spectral modulator and a gray scale sensor array to capture a multiplexed signal of the imaged scene. By designing the LC spectral modulator in compliance with the Compressive Sensing (CS) guidelines and applying appropriate algorithms we demonstrated reconstruction of spectral (hyper/ ultra) datacubes from an order of magnitude fewer samples than taken by conventional sensors. The LC modulator is designed to have an effective width of a few tens of micrometers, therefore it is prone to imperfections and spatial nonuniformity. In this work, we present the study of this nonuniformity and present a mathematical algorithm that allows the inference of the spectral transmission over the entire cell area from only a few calibration measurements.
Conference Presentation
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Liat Shecter, Liat Shecter, Yaniv Oiknine, Yaniv Oiknine, Isaac August, Isaac August, Adrian Stern, Adrian Stern, } "Method and algorithm for efficient calibration of compressive hyperspectral imaging system based on a liquid crystal retarder", Proc. SPIE 10410, Unconventional and Indirect Imaging, Image Reconstruction, and Wavefront Sensing 2017, 104100C (6 September 2017); doi: 10.1117/12.2278186; https://doi.org/10.1117/12.2278186
PROCEEDINGS
8 PAGES + PRESENTATION