Paper
14 February 2018 Full optical model of micro-endoscope with optical coherence microscopy, multiphoton microscopy and visible capabilities
David Vega, Kelli C. Kiekens, Nikolas C. Syson, Gabriella Romano, Tressa Baker, Jennifer K. Barton
Author Affiliations +
Proceedings Volume 10470, Endoscopic Microscopy XIII; 104700M (2018) https://doi.org/10.1117/12.2285059
Event: SPIE BiOS, 2018, San Francisco, California, United States
Abstract
While Optical Coherence Microscopy (OCM), Multiphoton Microscopy (MPM), and narrowband imaging are powerful imaging techniques that can be used to detect cancer, each imaging technique has limitations when used by itself. Combining them into an endoscope to work in synergy can help achieve high sensitivity and specificity for diagnosis at the point of care. Such complex endoscopes have an elevated risk of failure, and performing proper modelling ensures functionality and minimizes risk. We present full 2D and 3D models of a multimodality optical micro-endoscope to provide real-time detection of carcinomas, called a salpingoscope. The models evaluate the endoscope illumination and light collection capabilities of various modalities. The design features two optical paths with different numerical apertures (NA) through a single lens system with a scanning optical fiber. The dual path is achieved using dichroic coatings embedded in a triplet. A high NA optical path is designed to perform OCM and MPM while a low NA optical path is designed for the visible spectrum to navigate the endoscope to areas of interest and narrowband imaging. Different tests such as the reflectance profile of homogeneous epithelial tissue were performed to adjust the models properly. Light collection models for the different modalities were created and tested for efficiency. While it is challenging to evaluate the efficiency of multimodality endoscopes, the models ensure that the system is design for the expected light collection levels to provide detectable signal to work for the intended imaging.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David Vega, Kelli C. Kiekens, Nikolas C. Syson, Gabriella Romano, Tressa Baker, and Jennifer K. Barton "Full optical model of micro-endoscope with optical coherence microscopy, multiphoton microscopy and visible capabilities", Proc. SPIE 10470, Endoscopic Microscopy XIII, 104700M (14 February 2018); https://doi.org/10.1117/12.2285059
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissue optics

Navigation systems

Endoscopes

Sensors

Geometrical optics

3D modeling

Reflectivity

Back to Top