8 February 2018 Temperature increasing in titanium implants using a high-intensity diode laser for peri-implantitis decontamination
Author Affiliations +
Proceedings Volume 10473, Lasers in Dentistry XXIV; 104730P (2018) https://doi.org/10.1117/12.2289871
Event: SPIE BiOS, 2018, San Francisco, California, United States
Abstract
Peri-implantitis is a destructive inflammatory process that affects the tissues that provide support to the dental implant, the bone and gingiva, and can lead to the loss of the implant. Among the treatments of this disease, the irradiation of the contaminated surface with high intensity lasers is considered a promising alternative; Thus, irradiation parameters must be correctly adjusted in order to promote an efficient and safe treatment. This study investigated the temperature changes at the implant-bone interface during simulated implant surface decontamination using an 808nm diode laser. Dental implants were inserted in bovine bone, in which an artificial periimplant bone defect was made. Access holes of 0.5mm diameter were drilled to allow the positioning of four Ktype thermocouples in different regions: T0 Implant-bone interface, T1 inside the implant, T2 In the bone defect, T3 In the apex of the implant. For laser irradiation, an optical fiber was used at a distance of 0.5mm from the implant surface, and the mean output power varied between 0.5 to 3.0W on both pulsed (PW) and continuous (CW) wave modes. Irradiations were performed by 60s, and the temperature rises were registered for a period of 180s. It was observed that the critical threshold of 47ºC was exceeded at T0, T1 e T2 thermocouples when irradiations were performed at 1.0W; for T3 thermocouple, the threshold was exceeded at 3.0W CW mode. For PW mode, the thermocouples T0, T1, T2 had the threshold exceeded at the power of 1,0W and for T3 the threshold was exceeded at 3.0W. Decontamination of implant surfaces using the diode laser did not excessively heat the implant-bone interface within the mean output power ranging from 0.5 to 1.0W; however, the temperature rise is critical when using the mean power of 0.5W CW and 1.0W PW. Thus, using the PW mode up to the power of 1W seems to be a promising parameter for a safe clinical application.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
C. N. Faustino, A. P. Ana, and L. Bachman "Temperature increasing in titanium implants using a high-intensity diode laser for peri-implantitis decontamination", Proc. SPIE 10473, Lasers in Dentistry XXIV, 104730P (8 February 2018); https://doi.org/10.1117/12.2289871
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Bone

Semiconductor lasers

Tissues

Titanium

Laser irradiation

Natural surfaces

Continuous wave operation

Back to Top