9 March 2018 Imaging performance of CMOS and a-Si:H flat-panel detectors for C-arm fluoroscopy and cone-beam CT
Author Affiliations +
Abstract
Purpose: CMOS detectors are a potentially advantageous sensor technology for indirect-detection flat-panel detectors (FPDs), offering finer pixel pitch, faster frame rate, and lower electronic noise compared to a-Si:H sensors. This work presents a preliminary analysis of the 2D and 3D imaging performance of both detector technologies.

Methods: Two mobile C-arms were equipped with CMOS (Xineos 3030HS) and a-Si:H (PaxScan 3030X) FPDs. Technical assessment includes measurement of spatial resolution (MTF), image noise (NPS), and detective quantum efficiency (DQE). Evaluation of CBCT performance considers soft tissue visibility including axial image MTF, NPS, and noise-equivalent quanta (NEQ).

Results: The CMOS detector exhibited lower readout noise and slightly higher spatial resolution as expected. The a- Si:H detector showed about 10-15% higher DQE at low spatial frequencies while the CMOS detector showed greater resilience in DQE at higher spatial frequencies. In matched resolution CBCT, both detectors showed roughly equivalent performance.

Conclusion: CMOS detectors benefit performance with respect to high-frequency tasks, but the current work did not demonstrate strong advantage with respect to low-contrast soft-tissue visualization, in part due to light losses in scintillator-semiconductor coupling. Additional advantages include improved frame rate (reduced CBCT scan time). Ongoing work includes further investigation of modified bandwidth filters to take better advantage of underlying noiseresolution properties.
Conference Presentation
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
N. M. Sheth, M. W. Jacobson, W. Zbijewski, G. Kleinszig, S. Vogt, S. Soellradl, J. Bialkowski, W. S. Anderson, C. R. Weiss, G. M. Osgood, and J. H. Siewerdsen "Imaging performance of CMOS and a-Si:H flat-panel detectors for C-arm fluoroscopy and cone-beam CT", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105730M (9 March 2018); https://doi.org/10.1117/12.2293780
PROCEEDINGS
6 PAGES + PRESENTATION

SHARE
Advertisement
Advertisement
RELATED CONTENT

Image Quality In Selenium-Based Digital Radiography
Proceedings of SPIE (June 12 1986)
Resolution characteristics of afocal array optics
Proceedings of SPIE (November 15 2005)
Image Quality: Two Current Issues
Proceedings of SPIE (April 01 1987)
Correction of a liquid lens for 3D imaging systems
Proceedings of SPIE (April 27 2012)

Back to Top