9 March 2018 Spatio-energetic cross-talk in photon counting detectors: numerical detector model (PcTK) and workflow for CT image quality assessment
Author Affiliations +
Abstract
We developed and reported an analytical model (version 2.1) of inter-pixel cross-talk of energy-sensitive photon counting detectors (PCDs) in 2016 [1]. Since the time, we have identified four problems that are inherent to the design of the model version 2.1. In this study, we have developed a new model (version 3.2; “PcTK” for Photon Counting Toolkit) based on a totally different design concept. The comparison with the previous model version 2.1 and a Monte Carlo (MC) simulation has shown that the new model version 3.2 addressed the four problems successfully. The workflow script for computed tomography (CT) image quality assessment has demonstrated the utility of the model and potential values to CT community. The software packages including the workflow script, built using Matlab 2016a, has been made available to academic researchers free of charge (PcTK; https://pctk.jhu.edu).
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Katsuyuki Taguchi, Katsuyuki Taguchi, Karl Stierstorfer, Karl Stierstorfer, Christoph Polster, Christoph Polster, Okkyun Lee, Okkyun Lee, Steffen Kappler, Steffen Kappler, } "Spatio-energetic cross-talk in photon counting detectors: numerical detector model (PcTK) and workflow for CT image quality assessment", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 1057310 (9 March 2018); doi: 10.1117/12.2293881; https://doi.org/10.1117/12.2293881