9 March 2018 Scatter correction for multi-slice CT system
Author Affiliations +
Abstract
Scatter radiation in multi-slice CT system nowadays is playing a more and more important role, as the slice number is getting larger and larger, e.g. from 16 to 64 and even more. Scatter radiation may downgrade image quality e.g. create inhomogeneity and decrease the image contrast. Although in current multi-slice CT, anti-scatter collimation is widely used to reduce the scatter radiation received, as the detector is getting wider, its efficiency is getting weaker. Although beam hardening correction can somehow guarantee image homogeneity, as beam hardening effect and scatter radiation have different physical origin, a delegated scatter correction algorithm is desired. In this paper, we would like to propose a scatter correction algorithm working during data pre-processing before image reconstruction. After we implemented this algorithm in Siemens latest released Somatom Go. CT system, we obtain good image quality especially under certain clinical cases.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yang Wang, Yang Wang, Karl Stierstorfer, Karl Stierstorfer, Martin Petersilka, Martin Petersilka, Michael Grasruck, Michael Grasruck, Yi Tian, Yi Tian, } "Scatter correction for multi-slice CT system", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105733K (9 March 2018); doi: 10.1117/12.2293117; https://doi.org/10.1117/12.2293117