9 March 2018 Determination of optimal image type and lowest detectable concentration for iodine detection on a photon counting detector-based multi-energy CT system
Wei Zhou, Rachel Schornak, Gregory Michalak, Jayse Weaver, Dilbar Abdurakhimova, Andrea Ferrero, Kenneth A. Fetterly, Cynthia H. McCollough, Shuai Leng
Author Affiliations +
Abstract
Photon counting detector (PCD) based multi-energy CT is able to generate different types of images such as virtual monoenergetic images (VMIs) and material specific images (e.g., iodine maps) in addition to the conventional single energy images. The purpose of this study is to determine the image type that has optimal iodine detection and to determine the lowest detectable iodine concentration using a PCD-CT system. A 35 cm body phantom with iodine inserts of 4 concentrations and 2 sizes was scanned on a research PCD-CT system. For each iodine concentration, 80 repeated scans were performed and images were reconstructed for each energy threshold. In addition, VMIs at different keVs and iodine maps were also generated. CNR was measured for each type of images. A channelized Hotelling observer was used to assess iodine detectability after being validated with human observer studies, with area under the ROC curve (AUC) as a figure of merit. The agreement between model and human observer performance indicated that model observer could serve as an effective approach to determine optimal image type for the clinical practice and to determine the lowest detectable iodine concentration. Results demonstrated that for all size and concentration combinations, VMI at 70 keV had similar performance as that of threshold low images, both of which outperformed the iodine map images. At the AUC value of 0.8, iodine concentration as low as 0.2 mgI/cc could be detected for an 8 mm object and 0.5 mgI/cc for a 4 mm object with a 5 mm slice thickness.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Wei Zhou, Rachel Schornak, Gregory Michalak, Jayse Weaver, Dilbar Abdurakhimova, Andrea Ferrero, Kenneth A. Fetterly, Cynthia H. McCollough, and Shuai Leng "Determination of optimal image type and lowest detectable concentration for iodine detection on a photon counting detector-based multi-energy CT system", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105734U (9 March 2018); https://doi.org/10.1117/12.2294949
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Iodine

Computed tomography

Performance modeling

Photon counting

Sensors

X-ray computed tomography

Radiology

Back to Top