24 April 2018 Lateral spatial resolution improvement in laser scanning fluorescence microscopy using a subdiffraction limit optical spot
Author Affiliations +
Proceedings Volume 10711, Biomedical Imaging and Sensing Conference; 1071109 (2018) https://doi.org/10.1117/12.2319336
Event: SPIE Structured Light, 2018, Yokohama, Japan
Abstract
This paper presents lateral spatial resolution improvement by scanning a focused spot array pattern. To enhance the effective point spread function (PSF), we employed a phase-only computer generated hologram (CGH) that can reduce the spot size comparing to the single diffraction limited spot. A CGH was designed based on the Gerchberg-Saxton algorithm with a specific constraint to control the dispersion of light energy and the phase of generated spots. As a design example, we obtained a CGH for generating 3x3 optical spots whose sizes were reduced to 78.5% of the single diffraction limited spot. We also confirmed by simulation that the effective PSF was improved from 208 nm to 183 nm when using the subdiffraction limit spots for excitation.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Takahiro Nishimura, Daiki Shinkawa, Yusuke Ogura, Yosuke Tamada, Jun Tanida, "Lateral spatial resolution improvement in laser scanning fluorescence microscopy using a subdiffraction limit optical spot", Proc. SPIE 10711, Biomedical Imaging and Sensing Conference, 1071109 (24 April 2018); doi: 10.1117/12.2319336; https://doi.org/10.1117/12.2319336