24 April 2018 A 2.8-mm beam diameter system for retinal imaging with OCT and adaptive optics
Author Affiliations +
Proceedings Volume 10711, Biomedical Imaging and Sensing Conference; 107110C (2018) https://doi.org/10.1117/12.2319383
Event: SPIE Structured Light, 2018, Yokohama, Japan
Abstract
A less-complex 2.8-mm beam diameter spectral domain optical coherence tomography system with an adaptive optics module presented. In this system a Shack-Hartmann wavefront sensor used for aberration sensing and the Deformable mirror used for aberration correction. We demonstrated the diffraction-limited resolution performance of this system on model retina. On the model, measured speckle size with present system is 2.2 times smaller than a 1.2-mm beam diameter OCT system. Further, on the model eye SNR gain of 6.7 dB was quantified with the present system over a 1.2-mm beam diameter OCT system. Relatively small size of 25 cm by 50 cm, less complexity, large field of view of the present system as compared to the conventional AO-OCT systems, would make it suitable for ophthalmic clinical applications.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Maddipatla Reddikumar, Maddipatla Reddikumar, Joel Cervantes, Joel Cervantes, Yukitoshi Otani, Yukitoshi Otani, Barry Cense, Barry Cense, } "A 2.8-mm beam diameter system for retinal imaging with OCT and adaptive optics", Proc. SPIE 10711, Biomedical Imaging and Sensing Conference, 107110C (24 April 2018); doi: 10.1117/12.2319383; https://doi.org/10.1117/12.2319383