Paper
30 January 1990 Fiber-Optic Absorptive Oximeter
Andrzej W. Domanski, Stanislaw Kostrzewa, Tomasz R. Wolinski
Author Affiliations +
Proceedings Volume 1085, Optical Fibres and Their Applications V; (1990) https://doi.org/10.1117/12.952986
Event: Optical Fibers and Their Applications V, 1989, Warsaw, Poland
Abstract
In this paper the model of the absorptive fiber-optic oximeter with specially designed fiber-optic head coupled to two light-emitting diodes and a photodiode is presented. We constructed the blood chamber allowing the tips of optical fibers to be in contact with the blood under measurements. The absorptive fiber-optic oximeter was clinically tested in Central Hospital for Children in Warsaw. Results obtained were compared with the oxygen saturation measured by the classical oximeter OSM2 Hemoximeter.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrzej W. Domanski, Stanislaw Kostrzewa, and Tomasz R. Wolinski "Fiber-Optic Absorptive Oximeter", Proc. SPIE 1085, Optical Fibres and Their Applications V, (30 January 1990); https://doi.org/10.1117/12.952986
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Blood

Oximeters

Fiber optics

Optical fibers

Oxygen

Infrared radiation

Light emitting diodes

Back to Top