26 February 2019 Clinical evaluation of automated capillary refill time estimation in dogs and cats
Author Affiliations +
Abstract
In this study, we clinically evaluated a pulse oximeter-based device for automated capillary refill time (CRT) estimation in dogs and cats. CRT can reveal conditions like shock or anemia in dogs and cats. However, visual CRT estimation has low repeatability, and the available optical systems for automated estimation are not suitable for pets. We evaluated a custom-made portable CRT measuring device on various measurement sites of 12 dogs and 11 cats with parallel visual CRT estimation on the gum by treating veterinarian. The capillary refill was also recorded by a video camera for reference. The visual and video procedures were moderately correlated with the coefficient of 0.61; visual CRT values were on average for 0.18 s longer than the reference. On average, ~32% of measurements with the proposed device were successful. The rest failed due to excessive pigmentation, motion artifacts, and other pressure-induced effects. The measurement sites of the metacarpal pad, digit, and tail were moderately correlated with the reference values with coefficients of 0.53, 0.58, and 0.42, respectively.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Blaž Cugmas, Eva Štruc, and Janis Spigulis "Clinical evaluation of automated capillary refill time estimation in dogs and cats", Proc. SPIE 10868, Advanced Biomedical and Clinical Diagnostic and Surgical Guidance Systems XVII, 108681N (26 February 2019); https://doi.org/10.1117/12.2507620
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
CRTs

Visualization

Video

Capillaries

Tissues

Optical components

Skin

Back to Top