25 May 1989 Stereotactic Neurosurgery Planning On A PC Based Workstation.
Author Affiliations +
Abstract
Stereotactic surgery requires knowledge of cerebral structures derived from more than one image source. We have developed a PC-AT based workstation which accepts patient images, made with the stereotactic frame in place, from CT, MRI and DSA modalities. Reference markers on the frame are identified in the images to establish the coordinate geometry for each modality. Target points may be identified on each image type and trajectories of probe paths to these points defined. Targets identified on one set of images may be transferred automatically to other images of the same patient, in order, for example, to guarantee a vascular free path of approach to a target point deep within the brain. To date several hundred patients have had stereotactic surgery performed on the basis of plans using this system. Procedures included biopsy and aspiration of lesions, implantation of electrodes for the recording of deep EEG signals, and radiosurgical techniques based on the use of a high energy linear accelerator. We present clinical examples of the use of this system in typical stereotactic neurosurgery procedures, address stereoscopic applications, and discuss the results of inter-modality tests to establish the accuracy of the technique.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
T. M. Peters, J. A. Clark, G. B. Pike, C. Henri, L. Collins, D. Leksell, O. Jeppsson, "Stereotactic Neurosurgery Planning On A PC Based Workstation.", Proc. SPIE 1092, Medical Imaging III: Image Processing, (25 May 1989); doi: 10.1117/12.953298; https://doi.org/10.1117/12.953298
PROCEEDINGS
9 PAGES


SHARE
Back to Top