1 March 2019 Optimal acquisition setting for dual-contrast imaging with gadolinium and iodine on a research whole-body photon-counting-detector (PCD) CT
Author Affiliations +
Abstract
Photon-counting detectors (PCDs) can resolve the energy of incident x-ray photons, which allows simultaneous imaging of two contrast materials, such as iodine (I) and gadolinium (Gd), with a single scan. This capability may allow reduction of patient radiation dose for clinical applications that typically require multi-phase acquisitions by injecting different contrast media at different times and scanning only once to differentiate, for example, venous and arterial phases. The material decomposition performance on PCD-CT is dependent on acquisition setup including tube potential and energy thresholds. In this work, we performed a phantom study to evaluate the optimal acquisition settings for dual-contrast imaging using I and Gd on a research PCD-CT system. We further compared our results with a clinical dual-source dual-energy (DSDE) CT. An abdomen-shaped water phantom with I and Gd inserts of different concentrations was scanned using different energy thresholds and tube potentials to identify the optimal setup for I and Gd quantification. Results demonstrated that accurate quantification of I and Gd concentration was possible using the PCD-CT system. A tube potential of 80 kV and an energy threshold close to the K-edge of Gd (50 keV) was found to yield the best performance in terms of measurement root-mean-square-error (RMSE = 4.4 mg/mL for I and RMSE = 3.3 mg/mL for Gd). Further, the performance of PCD-CT with optimized setup was found to outperform DSDE-CT (RMSE = 8.1 mg/mL for I and 5.7 mg/mL for Gd).
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shengzhen Tao, Yizhong Wu, Kishore Rajendran, Cynthia H. McCollough, and Shuai Leng "Optimal acquisition setting for dual-contrast imaging with gadolinium and iodine on a research whole-body photon-counting-detector (PCD) CT", Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109484I (1 March 2019); https://doi.org/10.1117/12.2513454
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Gadolinium

X-ray computed tomography

Sensors

Computed tomography

Iodine

Imaging systems

Radiology

Back to Top