You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
27 March 2019Non-iterative method for metal artifact reduction by using a linearized beam-hardening correction model for polychromatic x-ray CT
The beam-hardening effect is one of the most important factors of metal artifact that degrades CT image quality. In the polychromatic X-ray, this occurs noticeably when scanning metallic materials with large changes in energy-dependent attenuation coefficient. This violates the assumption of a CT reconstruction based on a fixed attenuation coefficient in a monochromatic X-ray, which leads to beam-hardening artifacts such as streaking and cupping shapes. Numerous studies have been researched to reduce the beam-hardening artifacts. Most of the methods need the optimization based on iterative reconstruction, which causes a time-consuming problem. This study aims at an efficient methodology in terms of performance time while providing acceptable correction of beam-hardening artifacts. For this, the attenuation coefficient error due to beam hardening is modeled with respect to the length of the X-ray passing through the metallic material. And the model is approximated by a linear combination of four basis functions determined by the length. The linearity is also preserved in the reconstruction image, so that the coefficient of each basis function can be obtained by solving the minimization problem of the variance of the homogeneous metal region in the image. For the evaluation, a phantom including three titanium rods was scanned by a cone-beam CT system (Ray, South Korea) and the images were reconstructed by the standard FDK algorithm. The results showed that the proposed method is superior in terms of speed while delivering acceptable beam-hardening correction compared to recent methods. The proposed model will be effective for the applications where processing speed is important for the beam-hardening correction.
Jin Hur,Dongjoon Kim,Sangchul Lee, andYeong-gil Shin
"Non-iterative method for metal artifact reduction by using a linearized beam-hardening correction model for polychromatic x-ray CT
", Proc. SPIE 11050, International Forum on Medical Imaging in Asia 2019, 1105008 (27 March 2019); https://doi.org/10.1117/12.2521375