19 July 2019 Durable 3D test standards from mouse retina for performance testing of high-resolution optical coherence tomography systems
Author Affiliations +
Abstract
For performance testing of high-resolution label-free optical coherence tomography (OCT) systems, durable test standards with optical properties similar to native biological tissues are highly desirable. Therefore, we explored the establishment of ex-vivo 3D samples from dissected mouse retina. To achieve high stability, dissected tissue parts were embedded in EPON resin. Results from comparative investigations with OCT on murine retina in agarose demonstrate that by utilization of appropriate EPON embedding protocols durable samples with adequate refractive index properties can be created in which even tiny structures near the OCT resolution limits can be detected.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Álvaro Barroso, Peter Heiduschka, Gerburg Nettels-Hackert, Steffi Ketelhut, Björn Kemper, and Jürgen Schnekenburger "Durable 3D test standards from mouse retina for performance testing of high-resolution optical coherence tomography systems", Proc. SPIE 11078, Optical Coherence Imaging Techniques and Imaging in Scattering Media III, 110781V (19 July 2019); https://doi.org/10.1117/12.2527163