Abstract
Side lobe artifacts on point spread functions can be traced back to (1) fringe visibility variation across the spectrum, (2) errors in sampling instances, and (3) window functions. We demonstrate signal processing methods for correcting for all three of these issues. These methods require a system calibration step. If the systems slowly age, the recalibration step could be performed in the field with a fixtured target.
Conference Presentation
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bart Johnson, Tim N. Ford, Seungbum Woo, Walid Atia, Joey Jabbour, Nate Kemp, Muzammil A. Arain, Tilman Schmoll, Rick Williams, and Peter Whitney "Achieving the ideal point spread in swept source OCT", Proc. SPIE 11228, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIV, 112281H (21 February 2020); https://doi.org/10.1117/12.2543082