27 November 1989 A 1024 X 1024 CCD For Scientific Applications
Author Affiliations +
Abstract
In order to satisfy the increasing interest in the large format scientific grade CCD with high resolution capability, a full frame 1024 x 1024 four phase CCD has been designed, fabricated and tested. The devices show a 5 electrons noise at - 100°C and very good transfer efficiency. A four quadrant organization with four video outputs allows a 20 ms minimum readout time. Performances at room temperature and low temperature will be presented. Plan for future enhancements -back side illumination and high resistivity silicon substrate- will be discussed.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gerard Beal, Jean-Alain Cortiula, Marie-Noelle Gaujour, Gilles Boucharlat, Jean Chabbal, Alain Borestel, "A 1024 X 1024 CCD For Scientific Applications", Proc. SPIE 1159, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy and Atomic Physics, (27 November 1989); doi: 10.1117/12.962593; https://doi.org/10.1117/12.962593
PROCEEDINGS
7 PAGES