15 February 2021 Back-end readout electronic design and initial results: a head-and-neck dedicated PET system based on CZT
Author Affiliations +
Abstract
Cadmium zinc telluride (CZT) radiation detectors are suitable for various applications, due to good energy performance at room temperature and simple pixilation to achieve high spatial resolution. Our group is developing a two-panel head-and-neck dedicated positron emission tomography (PET) system with CZT detectors. In this work, we present the back-end readout electronic design and the initial electronic noise results of our system. The back-end readout electronic incorporates RENA boards (a total 150 RENA boards for each panel), 30:1 fan-in boards connecting 30 RENA boards to the PicoZed 7010/7020 board (a total 5 fan-in boards for each panel). In each panel, 5:1 intermediate boards are used for biasing CZT detectors. The RENA board and the Picozed board are capable of data transmission of 50 Mbps and 6.6 Gbps, respectively. Electronic noise was also quantified using a square wave test pulse that provides charge injection into each channel of the RENA chip in the amount of 75fC/Volt. The pulse amplitude was chosen to generate approximately the same amount of charges as 511 keV photon would provide for each channels. The FWHM electronic noise at 511 keV was measured to be less than 1% (FWHM of 7.80 ± 1.47 ADC units or 4.89 ± 0.92 keV after conversion).
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yuli Wang, Ryan Herbst, and Shiva Abbaszadeh "Back-end readout electronic design and initial results: a head-and-neck dedicated PET system based on CZT", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 1159510 (15 February 2021); https://doi.org/10.1117/12.2576598
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Positron emission tomography

Sensors

Cadmium

Data transmission

Spatial resolution

Zinc

Back to Top