15 February 2021 Comparison of skin dose calculated by the dose tracking system (DTS) with a beam angular correction factor and that calculated by Monte-Carlo
Author Affiliations +
Abstract
Skin dose is dependent on the incident beam angle and corrections are needed for accurate estimation of the risk of deterministic effects of the skin. Angular-correction factors (ACF) were calculated and incorporated into our skin-dosetracking system (DTS) and the results compared to Monte-Carlo simulations for a neuro-interventional procedure. To obtain the ACF’s, EGSnrc Monte-Carlo (MC) software was used to calculate the dose averaged over 0.5, 1, 2, 3, 4 and 5 mm depth into the entrance surface of a water phantom at the center of the field as a function of incident beam to skin angle from 90-10 degrees for beam field sizes from 5-15 cm and for beam energies from 60-120 kVp. These values were normalized to the incident primary dose to obtain the ACF. The angle of incidence at each mesh vertex in the beam on the surface of the DTS patient graphic was calculated as the complement of the angle between the normal vector and the vector of the intersecting ray from the tube focal spot; skin dose at that vertex was calculated using the corresponding ACF. The skin-dose values with angular correction were compared to those calculated using MC with a matching voxelized phantom. The results show the ACF decreases with decreasing incident angle and skin thickness, and increases with increasing field size and kVp. Good agreement was obtained between the skin dose calculated by the angular-corrected DTS and MC, while use of the ACF allows the real-time performance of the DTS to be maintained.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sheng-Hsuan Sun, Stephen Rudin, and Daniel R. Bednarek "Comparison of skin dose calculated by the dose tracking system (DTS) with a beam angular correction factor and that calculated by Monte-Carlo", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 1159522 (15 February 2021); https://doi.org/10.1117/12.2580938
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Skin

Monte Carlo methods

Visualization

Back to Top