You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
1 July 1990Physical and psychophysical analysis of laser readout conditions for imaging plates in computed radiography
Computed radiography uses a photostimulable phosphor coated imaging plate which is
exposed to x-rays and laser read to form an image. After laser reading, there is a
considerable amount of energy remaining on the imaging plate which is not used. This study
simulated a change in the laser readout process to utilize more of the energy on the image
plate, and potentially improve image quality without changing exposure factors. Images of a
contrast detail phantom were made before and after alteration of the readout process and
analyzed by both physical and psychophysical means. It was found that there is an increase
in the signal-to-noise ratio, when measured with an aperture of the size of a single pixel
(linear dimension about 0.15 mm). However there is no change in the signal-to-noise ratio,
when measured with apertures of the size of 0.75 mm (5 x 5 pixels) and 1.5 mm (10 x 10
pixels). This agrees with the results of the contrast detail study: after alteration, the
observers did not detect smaller objects than they had before the alteration. It appears the
imaging plate readout process is fairly optimized.
The alert did not successfully save. Please try again later.
Brigid Gerety, Hans Roehrig, George W. Seeley, Tinglan Ji, David A. Yocky, Bryan Westerman, Theron W. Ovitt, "Physical and psychophysical analysis of laser readout conditions for imaging plates in computed radiography," Proc. SPIE 1232, Medical Imaging IV: Image Capture and Display, (1 July 1990); https://doi.org/10.1117/12.18839