Paper
1 July 1990 Final tests and commissioning of the UCL echelle spectrograph
Francisco Diego, Andrew Charalambous, Adrian C. Fish, David D. Walker
Author Affiliations +
Abstract
The optical design and performance of the UCLES echelle spectrograph, installed at the f/37.7 coude focus of the 3.9-m AAT in June 1988, are described and illustrated with extensive diagrams, drawings, photographs, and sample spectra. The UCLES operates at 300-1100 nm with resolution 30,000-115,000 and adjustable collimated beam size; it employs either 31.6-g/mm or 79-g/mm echelle gratings and a train of UV-transmitting fused-silica prisms for cross-dispersion. Also discussed are the focal modifier lenses; the Bowen-Walraven image slicer; the commissioning procedures; and preliminary observations of Zeta Oph, planetary nebulae, and Seyfert galaxies.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Francisco Diego, Andrew Charalambous, Adrian C. Fish, and David D. Walker "Final tests and commissioning of the UCL echelle spectrograph", Proc. SPIE 1235, Instrumentation in Astronomy VII, (1 July 1990); https://doi.org/10.1117/12.19119
Lens.org Logo
CITATIONS
Cited by 41 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Prisms

Astronomy

Cameras

Lenses

Calibration

Sensors

Charge-coupled devices

RELATED CONTENT