Abstract
The effect of different interpolation methods on imaging quality of swept-source optical coherence tomography (SSOCT) is analyzed. A 1310nm SS-OCT system is built for imaging. Time-domain interpolation method based on spectral phase is used for resampling. The interpolation methods include piecewise linear interpolation, piecewise quadratic parabola interpolation, piecewise cubic Hermite interpolation and piecewise cubic spline interpolation. The image of the plane mirror, human skin and eyes are compared by the above four interpolation methods. Experimental results demonstrate that the piecewise cubic spline interpolation has the best image quality, which is the smoothest and has the lowest noise.
© (2022) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xiao Liang, Heping Li, and Pinghe Wang "Effects of different interpolation methods on imaging quality of (SS-OCT) system", Proc. SPIE 12478, Thirteenth International Conference on Information Optics and Photonics (CIOP 2022), 124780X (15 December 2022); https://doi.org/10.1117/12.2651726
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Imaging systems

Image segmentation

Back to Top