11 August 2023 Dual ultrahigh speed swept-source and time domain optical coherence tomography system using a time stretch laser and a KTN deflector
Author Affiliations +
Abstract
We present an ultrahigh-speed Swept Source Optical Coherence Tomography (SS-OCT) system that allows a volume rate of 400 Hz paired with a Time Domain (TD) subsystem. For the SS-OCT, a 40 MHz swept source is used, while for the TD-OCT, a broadband source. Both systems employ a scanning system that consists of a KTN scanner paired with a galvoscanner. The KTN crystal scans the beam laterally at 100 kHz. This allows B-scan OCT repetition rate, while with the galvoscanner 200 lines are obtained at 400 Hz in the en-face display of both SS-OCT and TD-OCT systems.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alejandro Martínez Jiménez, Sacha Grelet, Patrick Bowen Montague, Adrian Bradu, and Adrian Podoleanu "Dual ultrahigh speed swept-source and time domain optical coherence tomography system using a time stretch laser and a KTN deflector", Proc. SPIE 12632, Optical Coherence Imaging Techniques and Imaging in Scattering Media V, 126320B (11 August 2023); https://doi.org/10.1117/12.2670440